For radiopharmaceutical quality control
Tec-Control Chromatography tests the radiochemical
purity of specific Tc-99m-labeled radiopharmaceuticals. The
accompanying chart shows which strips and solvents are
required to perform each individual test. Some solvents must be
purchased separately
(see Sigma-Aldrich chart) due to hazardous material shipping
restrictions.
Detailed instruction manuals are packaged with each
strip container, although our
Radiopharmaceutical QC Procedure Manual (151-000) explains paper
chromatography
in greater detail.
TEC-CONTROL ALUMINUM BREAKTHRU
KIT
Simple 3-step procedure
The Aluminum Breakthru Kit provides a rapid, easy
and inexpensive way to test aqueous
solutions, particularly pertechnetate generator eluate, for trace
quantities of aluminum.
Aluminum forms an intense red precipitate with the indicator paper,
and the intensity of the
color is directly proportional to the amount of aluminum in the
solution. The USP allows a
concentration of aluminum ion in an injection ≤10 micrograms per
milliliter (10 µg/ml) in
technetium 99m eluate prepared from Molybdenum 99 formed as a result
of uranium fission.
Simple Procedure:
| 1. | Place a drop of the
eluate or solution to be tested on the indicator paper.
The best procedure is to form a hanging drop using a 19-22G needle. |
| 2. | Place a drop of the
standard aluminum solution on the indicator paper. Use the same size drop. |
| 3. | Compare the intensity of
the red spot formed. If the eluate spot is less intense
than the standard solution, the eluate contains less than 10µg/ml aluminum. |
|
|
|||
|
|||
|
|
|||
|
Miniaturized Chromatographic Quality-Control
Procedures for Tc-99m Radiopharmaceuticals;
A. Michael Zimmer and Dan G. Pavel, Journal of Nuclear Medicine,
Vol. 18/12, Dec. 1977, pg. 1230.
Technical Parameters Associated with Miniaturized Chromatography
Systems; Raimund A. Taukulis,
A. Michael Zimmer, Dan G. Pavel and Bhupendra A. Patel, University
of Illinois Medical Center,
Chicago, Illinois, Journal of Nuclear Medicine Technology, Vol. 7/1.
|
| * See Sigma-Aldrich product matrix for solvents. | |
| ** Note: A Well Plate is suggested to perform QC on these radiopharmacuticals | |
| Contact Fisher Scientific Corp. 800-766-7000 / www.fishersci.com / Item #14-245-71 | |
|
Tec-Control Solvent Vendor: Sigma-Aldrich Chemical Company 800-558-9160 / www.sigmaaldrich.com Note: Customers outside the US should visit Sigma-Aldrich web site to locate a regional office. |
|
| Solvent Description | Vendor Part # |
| Acetone HPLC Grade | 27072-5 |
| Ethyl Acetate 99.5% ACS Reagent | 31990-2 |
| Chloroform 99.8% ACS Reagent | 31998-8 |
| Tetrahydrofuran 99+% ACS Reagent | 36058-9 |



Radiopharmaceutical
QC Manual
This 90-page manual explains Paper
Chromatography, a QC method for evaluating the radiochemical purity
of currently used Tc-99m-labeled radiopharmaceuticals. Procedures
are quick and easy to use, a simple quality control solution for any
nuclear medicine department.